Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.

Recall number
Z-1703-2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.
Status
Terminated
Initiated
2015-03-25
Posted
2015-06-02
Terminated
2017-01-24
Reason
Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed. Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.
Root cause
Under Investigation by firm
Action
On March 24, 2015 an Urgent Medical Device Recall letter was mailed to all affected customers to notify them about the issue, associated risks, and actions to be taken.
Quantity
27501
Distribution
Worldwide Distribution. US naionwide including Puerto Rico, Barbados and Mexico.
Lot, serial or model codes
Catalog Number B4218-1, Material Number 10445709, lot numbers 557128, 557129, 557130, 557131, 557132, 557133, 557134, 557135, 557140, 557141 , expiration dates:10/31/2015, 11/21/2015, 2/16/2016
510(k) numbers
["K760318"]
PMA numbers
not on file
Product code
GFO
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28