FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0544CJDA
- Recall number
- Z-1701-2024
- Recalling firm
- Olympus Corporation of the Americas
- Product
- POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0544CJDA
- Status
- Open, Classified
- Initiated
- 2024-03-07
- Posted
- 2024-04-30
- Terminated
- not on file
- Reason
- Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
- Root cause
- Process control
- Action
- Olympus issued Urgent Medical Device Removal Letter on April 2, 2024 via email/visit. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory and identify any POWERSEAL products having the lot number identified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. Quarantine and cease use of the affected model/lots. 2. Please contact Customer Service at at 1-800-848-9024, option 2, with the quantity, model and lot number of the affected device(s). Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of the affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0444 . c. Complete the form as instructed and include your account ID number. 4. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@olympus.com
- Quantity
- 150 units: (20 units US); (130 units OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.
- Lot, serial or model codes
- UDI-DI:00821925044579 Lot Code: CA179370
- 510(k) numbers
- ["K212643"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28