Skip to the content

FDA Medical Device Recalls (openFDA)

Micro Therapeutics Inc, Dba Ev3 Neurovascular: MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g) REF 300016 (h) REF 300017 (i) REF 300018 Intended to restore blood flow by removing thrombus from a large intracranial vessel experiencing ischemic stroke within 8 hours of symptom onset.

Recall number
Z-1701-2018
Recalling firm
Micro Therapeutics Inc, Dba Ev3 Neurovascular
Product
MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g) REF 300016 (h) REF 300017 (i) REF 300018 Intended to restore blood flow by removing thrombus from a large intracranial vessel experiencing ischemic stroke within 8 hours of symptom onset.
Status
Terminated
Initiated
2018-02-26
Posted
2018-05-16
Terminated
2021-09-30
Reason
There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.
Root cause
Process design
Action
The firm initiated their recall via an Urgent Medical Device Recall letter, dated February 26, 2018. The letter identified the affected device, described the issue, product scope, and the actions to be taken. The firm recommended users remove and quarantine all affected and unexpired product in their inventory and return the affected products to Medtronic. Customers may contact their Medtronic representative if alternative product is needed. The Customer Confirmation Form should be completed and faxed to Medtronic at 1-949-434-5020, to the attention of Neurovascular Quality.
Quantity
1,942 units
Distribution
Worldwide Distribution
Lot, serial or model codes
All lot numbers: Product Number: UDI (a) REF 300010: 00843889031334, 00843889031341, 00847536020207 (b) REF 300011: 00843889031365, 00843889031372, 00847536020214 (c) REF 300012: 00843889031396, 00843889031402, 00847536020221 (d) REF 300013: 00843889031426, 00843889031433, 00847536020238 (e) REF 300014: NOT DISTRIBUTED (f) REF 300015: 00847536020122, 00847536020160 (g) REF 300016: 00847536020139, 00847536020177 (h) REF 300017: 00847536020146, 00847536020184 (i) REF 300018: 00847536020153, 00847536020191
510(k) numbers
["K141516"]
PMA numbers
not on file
Product code
NRY
Firm FEI
3004904811
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register