Skip to the content

FDA Medical Device Recalls (openFDA)

Angiodynamics Inc. (Navilyst Medical Inc.): Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.

Recall number
Z-1701-2016
Recalling firm
Angiodynamics Inc. (Navilyst Medical Inc.)
Product
Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.
Status
Terminated
Initiated
2016-03-21
Posted
2016-05-20
Terminated
2017-09-18
Reason
Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.
Root cause
Environmental control
Action
On 3/21/2016, AngioDynamics sent recall notification packages (dated 3/21/2016) to the consignees via Federal Express. Consignees are instructed to follow the instructions provided on the Maquet Cardiopulmonary recall letter (dated 2/23/2016) and return the Reply Verification Tracking Form, provided in the recall notification.
Quantity
Domestic: 339 kits
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot Numbers 37FE2901, 37GD2908, 37KD2405
510(k) numbers
["K092486"]
PMA numbers
not on file
Product code
DWE
Firm FEI
3017892510
Firm city
Glens Falls
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register