Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics: IMMULITE 2000 --- Insulin --- Catalog Number: L2KIN2 (200 test), L2KIN6 (600 tests) --- Test Code: INS, Color: Orange --- Product Usage: For in vitro diagnostic use with IMMULITE and IMMULITE 1000 Systems Analyzers --- for the quantitative measurement of insulin in serum or heparinized plasma, for the management of diabetes.

Recall number
Z-1701-2012
Recalling firm
Siemens Healthcare Diagnostics
Product
IMMULITE 2000 --- Insulin --- Catalog Number: L2KIN2 (200 test), L2KIN6 (600 tests) --- Test Code: INS, Color: Orange --- Product Usage: For in vitro diagnostic use with IMMULITE and IMMULITE 1000 Systems Analyzers --- for the quantitative measurement of insulin in serum or heparinized plasma, for the management of diabetes.
Status
Terminated
Initiated
2012-04-30
Posted
2012-06-02
Terminated
2017-03-16
Reason
Siemens Healthcare Diagnostics has observed a negative bias in the Immulite Insulin assay.
Root cause
Component change control
Action
Siemens sent an Urgent Device Recall letter dated April 2012 via fax to affected U.S. customers and an Urgent Field Safety Notice to customers outside the U.S. The letter identified the affected kit lots, reason for voluntary recall, risk to health and actions to be taken by the customer/user. Customers were instructed to discard any inventory of the affected kit lots and complete and return the attached Field Correction Effectiveness Check questionnaire within thirty (30) days to request a no-charge replacement kits. Customers were requested to post this field correction by each IMMULITE system to alert users to the situation. For questions contact your Siemens Technical Solutions Center.
Quantity
not on file
Distribution
Worldwide Distribution - US (nationwide)
Lot, serial or model codes
kit lots 312, 313, 314, 315, 316
510(k) numbers
["K022603"]
PMA numbers
not on file
Product code
CFP
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register