Waldemar Link GmbH & Co. KG (Mfg Site): Posterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2 - cemented lg), (5) 880-313/11 (Sz. 3-4 - cemented sm), (6) 880-313/12 (Sz. 3-4 - cemented md), (7) 880-313/21 (Sz. 3-4 - cemented lg), (8) 880-313/22 (Sz. 3-4 - cemented x-lg), (9) 880-315/11 (Sz. 5-6 - cemented sm), (10) 880-315/12 (Sz. 5-6 - cemented md), (11) 880-315/21 (Sz. 5-6 - cemented lg), (12) 880-315/22 (Sz. 5-6 - cemented x-lg), (13) 880-317/11 (Sz. 7-8 - cemented sm), (14) 880-317/12 (Sz. 7-8 - cemented md), (15) 880-317/21 (Sz. 7-8 - cemented lg), (16) 880-317/22 (Sz. 7-8 - cemented x-lg), (17) 880-319/11 (Sz. 9-10 - cemented sm), (18) 880-319/12 (Sz. 9-10 - cemented md), (19) 880-319/21 (Sz. 9-10 - cemented lg), (20) 880-319/22 (Sz. 9-10 - cemented x-lg).
- Recall number
- Z-1700-2025
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Product
- Posterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2 - cemented lg), (5) 880-313/11 (Sz. 3-4 - cemented sm), (6) 880-313/12 (Sz. 3-4 - cemented md), (7) 880-313/21 (Sz. 3-4 - cemented lg), (8) 880-313/22 (Sz. 3-4 - cemented x-lg), (9) 880-315/11 (Sz. 5-6 - cemented sm), (10) 880-315/12 (Sz. 5-6 - cemented md), (11) 880-315/21 (Sz. 5-6 - cemented lg), (12) 880-315/22 (Sz. 5-6 - cemented x-lg), (13) 880-317/11 (Sz. 7-8 - cemented sm), (14) 880-317/12 (Sz. 7-8 - cemented md), (15) 880-317/21 (Sz. 7-8 - cemented lg), (16) 880-317/22 (Sz. 7-8 - cemented x-lg), (17) 880-319/11 (Sz. 9-10 - cemented sm), (18) 880-319/12 (Sz. 9-10 - cemented md), (19) 880-319/21 (Sz. 9-10 - cemented lg), (20) 880-319/22 (Sz. 9-10 - cemented x-lg).
- Status
- Open, Classified
- Initiated
- 2025-04-03
- Posted
- 2025-04-30
- Terminated
- not on file
- Reason
- The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
- Root cause
- Nonconforming Material/Component
- Action
- On April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers. Actions to be taken by the Customer/User: Please take note of the following instructions for replacing any screws found missing the thread. No product return is required. Instructions for disassembling and replacing a defective screw: Disassembly 1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment. 2. Exchange the screw. Open an alternative Augment and use its screw. Alternative sizes of Augments are always available in the operating room during the surgery. For all Femoral Augment sizes and types the same long screw type is preassembled and for all Tibial Augments the same short screw type is preassembled, see table 1. Assembly/Replacement 3. Use the 2.5 mm Torque Wrench article REF (15-2545) to screw in the replacement screw into the Augment. 4. The screw head must be countersunk in the Augment and the screw tip must protrude. Follow the LinkSymphoKnee surgical technique for all other steps. Please respond using the Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the instructions provided above to replace any affected screws. Type of Action by the Company: LINK will provide sterilized replacement screws (for both Femoral Augments and Tibial Augments) for each affected distributor as a back-up for upcoming surgeries. Other Information: Opening an additional Augment to exchange the screw will not incur any costs to you. Should you have any questions about acquiring backup/replacement screws for forthcoming surgeries (once available), please contact Customer Service at LinkBio (order@linkbio.com). Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any additional questions.
- Quantity
- 908 units
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL, KS, LA, MD, MO, NV, OH, TX, VA, WI.
- Lot, serial or model codes
- Item Number/UDI-DI: (1) 880-310/11 (04026575257539), (2) 880-310/21 (04026575257546), (3) 880-311/11 (04026575257607), (4) 880-311/21 (04026575257584), (5) 880-313/11 (04026575257621), (6) 880-313/12 (04026575257645), (7) 880-313/21 (04026575257669), (8) 880-313/22 (04026575257690), (9) 880-315/11 (04026575257768), (10) 880-315/12 (04026575257775), (11) 880-315/21 (04026575257782), (12) 880-315/22 (04026575257799), (13) 880-317/11 (04026575257843), (14) 880-317/12 (04026575257850), (15) 880-317/21 (04026575257867), (16) 880-317/22 (04026575257881), (17) 880-319/11 (04026575257942), (18) 880-319/12 (04026575257959), (19) 880-319/21 (04026575257966), (20) 880-319/22 (04026575257973). Applies to all lots of the LINK SymphoKnee Femoral Augments listed, manufactured up until 03/01/2025.
- 510(k) numbers
- ["K202924"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 3003386935
- Firm city
- Norderstedt
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28