Waldemar Link GmbH & Co. KG (Mfg Site): Posterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2 - cemented lg), (5) 880-313/11 (Sz. 3-4 - cemented sm), (6) 880-313/12 (Sz. 3-4 - cemented md), (7) 880-313/21 (Sz. 3-4 - cemented lg), (8) 880-313/22 (Sz. 3-4 - cemented x-lg), (9) 880-315/11 (Sz. 5-6 - cemented sm), (10) 880-315/12 (Sz. 5-6 - cemented md), (11) 880-315/21 (Sz. 5-6 - cemented lg), (12) 880-315/22 (Sz. 5-6 - cemented x-lg), (13) 880-317/11 (Sz. 7-8 - cemented sm), (14) 880-317/12 (Sz. 7-8 - cemented md), (15) 880-317/21 (Sz. 7-8 - cemented lg), (16) 880-317/22 (Sz. 7-8 - cemented x-lg), (17) 880-319/11 (Sz. 9-10 - cemented sm), (18) 880-319/12 (Sz. 9-10 - cemented md), (19) 880-319/21 (Sz. 9-10 - cemented lg), (20) 880-319/22 (Sz. 9-10 - cemented x-lg).

Recall number
Z-1700-2025
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
Posterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2 - cemented lg), (5) 880-313/11 (Sz. 3-4 - cemented sm), (6) 880-313/12 (Sz. 3-4 - cemented md), (7) 880-313/21 (Sz. 3-4 - cemented lg), (8) 880-313/22 (Sz. 3-4 - cemented x-lg), (9) 880-315/11 (Sz. 5-6 - cemented sm), (10) 880-315/12 (Sz. 5-6 - cemented md), (11) 880-315/21 (Sz. 5-6 - cemented lg), (12) 880-315/22 (Sz. 5-6 - cemented x-lg), (13) 880-317/11 (Sz. 7-8 - cemented sm), (14) 880-317/12 (Sz. 7-8 - cemented md), (15) 880-317/21 (Sz. 7-8 - cemented lg), (16) 880-317/22 (Sz. 7-8 - cemented x-lg), (17) 880-319/11 (Sz. 9-10 - cemented sm), (18) 880-319/12 (Sz. 9-10 - cemented md), (19) 880-319/21 (Sz. 9-10 - cemented lg), (20) 880-319/22 (Sz. 9-10 - cemented x-lg).
Status
Open, Classified
Initiated
2025-04-03
Posted
2025-04-30
Terminated
not on file
Reason
The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Root cause
Nonconforming Material/Component
Action
On April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers. Actions to be taken by the Customer/User: Please take note of the following instructions for replacing any screws found missing the thread. No product return is required. Instructions for disassembling and replacing a defective screw: Disassembly 1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment. 2. Exchange the screw. Open an alternative Augment and use its screw. Alternative sizes of Augments are always available in the operating room during the surgery. For all Femoral Augment sizes and types the same long screw type is preassembled and for all Tibial Augments the same short screw type is preassembled, see table 1. Assembly/Replacement 3. Use the 2.5 mm Torque Wrench article REF (15-2545) to screw in the replacement screw into the Augment. 4. The screw head must be countersunk in the Augment and the screw tip must protrude. Follow the LinkSymphoKnee surgical technique for all other steps. Please respond using the Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the instructions provided above to replace any affected screws. Type of Action by the Company: LINK will provide sterilized replacement screws (for both Femoral Augments and Tibial Augments) for each affected distributor as a back-up for upcoming surgeries. Other Information: Opening an additional Augment to exchange the screw will not incur any costs to you. Should you have any questions about acquiring backup/replacement screws for forthcoming surgeries (once available), please contact Customer Service at LinkBio (order@linkbio.com). Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any additional questions.
Quantity
908 units
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL, KS, LA, MD, MO, NV, OH, TX, VA, WI.
Lot, serial or model codes
Item Number/UDI-DI: (1) 880-310/11 (04026575257539), (2) 880-310/21 (04026575257546), (3) 880-311/11 (04026575257607), (4) 880-311/21 (04026575257584), (5) 880-313/11 (04026575257621), (6) 880-313/12 (04026575257645), (7) 880-313/21 (04026575257669), (8) 880-313/22 (04026575257690), (9) 880-315/11 (04026575257768), (10) 880-315/12 (04026575257775), (11) 880-315/21 (04026575257782), (12) 880-315/22 (04026575257799), (13) 880-317/11 (04026575257843), (14) 880-317/12 (04026575257850), (15) 880-317/21 (04026575257867), (16) 880-317/22 (04026575257881), (17) 880-319/11 (04026575257942), (18) 880-319/12 (04026575257959), (19) 880-319/21 (04026575257966), (20) 880-319/22 (04026575257973). Applies to all lots of the LINK SymphoKnee Femoral Augments listed, manufactured up until 03/01/2025.
510(k) numbers
["K202924"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register