Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Syngo.plaza software Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Recall number
Z-1700-2018
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo.plaza software Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Status
Terminated
Initiated
2018-02-20
Posted
not on file
Terminated
2019-07-16
Reason
Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis.
Root cause
Software design
Action
On February 25, 2018 a Customer Information Letter was mailed to customers. This notification informed customers of the software changes related to All Sygno.plaza systems with VB20A. Customers were asked to install an updated software patch and instructed to share the notification with all users of the affected product. Customers with questions should call Siemens Customer Care Centre at (866)823-9702. For further questions, please call (610) 448-6461.
Quantity
92
Distribution
USA (nationwide)
Lot, serial or model codes
All Serial Numbers
510(k) numbers
["K132532"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register