Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Communications: The Stryker Visum LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures with high intensity light. Affected part numbers are 0682001432 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P/FP), 0682001286 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P/FP), 0682001298 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P), 0682001299 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P/5P), 0682001300 (EDS LIGHT SUSPENSION, CENTRAL AXIS, FP/5P/5P), 0682001433 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P), 0682001434 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P/5P), 0682001435 (EDS LIGHT SUSPENSION, CENTRAL AXIS, FP/9P/5P) Product Usage: The Stryker Visum LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures with high intensity light

Recall number
Z-1699-2017
Recalling firm
Stryker Communications
Product
The Stryker Visum LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures with high intensity light. Affected part numbers are 0682001432 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P/FP), 0682001286 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P/FP), 0682001298 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P), 0682001299 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P/5P), 0682001300 (EDS LIGHT SUSPENSION, CENTRAL AXIS, FP/5P/5P), 0682001433 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P), 0682001434 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P/5P), 0682001435 (EDS LIGHT SUSPENSION, CENTRAL AXIS, FP/9P/5P) Product Usage: The Stryker Visum LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures with high intensity light
Status
Terminated
Initiated
2017-02-28
Posted
2017-03-10
Terminated
2018-01-08
Reason
The potential hazard that can arise for the combination of the missing screws and the overloading by the end user is insufficient mount force that causes the equipment to fall.
Root cause
Device Design
Action
Stryker sent an Urgent Medical Device Correction letter dated March 3, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Distributors will notify customers via tracked Customer Notification Letter. A method of written communication will include a confirmation of delivery. For questions contact Stryker Technical Support at 800-243-5135 or comm.techservices@stryker.com
Quantity
292 devices (in US)
Distribution
Worldwide distribution - United States Nationwide - Australia, Canada, Great Britain, Japan, Netherlands and Puerto Rico
Lot, serial or model codes
"Serial Numbers:  100000000098399 100000000099187 100000000099193 100000000093937 100000000094391 100000000093936 100000000087555 100000000091032 100000000093938 100000000093928 100000000093939 100000000098397 100000000092672 100000000093927 100000000106675 100000000106685 100000000109560 100000000109569 100000000109570 100000000086360 100000000089194 100000000091766 100000000091778 100000000098401 100000000099325 100000000103228 100000000090520 100000000102434 100000000102439 100000000103111 100000000088303 100000000089467 100000000094993 100000000095927 100000000101580 100000000092699 100000000101590 100000000101591 100000000103686 100000000109291 100000000097048 100000000112633 100000000087549 100000000094982 100000000091762 100000000112607 100000000112609 100000000091531 100000000106681 100000000106682 100000000106684 100000000106159 100000000091728 100000000103271 100000000090522 100000000099356 100000000089970 100000000090530
510(k) numbers
["K060802"]
PMA numbers
not on file
Product code
FSY
Firm FEI
3002975342
Firm city
Flower Mound
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register