Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Incorporated: CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS

Recall number
Z-1696-2019
Recalling firm
Accuray Incorporated
Product
CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS
Status
Terminated
Initiated
2018-04-10
Posted
not on file
Terminated
2021-04-01
Reason
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Root cause
Other
Action
The firm issued a field safety notice by letter on 04/10/2019. The letter explained the issue and provided safety instructions. All impacted customers with affected systems will be contacted to arrange for a software upgrade.
Quantity
68 units
Distribution
US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
Lot, serial or model codes
a) 0660000: CyberKnife Treatment Delivery System Serial Numbers: C0434, C0436, C0437, C0438, C0440, C0441, C0443, C0444, C0445, C0446, C0447, C0448, C0449, C0450, C0453, C0454, C0456, C0457, C0458, C0460, C0464, C0465, C0466, C0472, C0475   b) 053101-010: CyberKnife M6 FI Treatment Delivery System Serial Numbers: C0334, C0381  c) 053201-010: CyberKnife M6 FM Treatment Delivery System Serial Numbers: C0363  d) 053301-010: CyberKnife M6 FIM Treatment Delivery System Serial Numbers: C0337, C0343, C0349, C0352, C0358, C0370, C0371, C0378, C0384, C0387, C0389, C0390, C0391, C0393, C0409, C0424, C0426, C0427, C0428, C0433, C0439   e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System Serial Numbers: C0330, C0332, C0342, C0344, C0345, C0348, C0359, C0360, C0364, C0367, C0372, C0394, C0398, C0400, C0402, C0404, C0415, C0419   f) 053501-010: CyberKnife Matrix Tier F TDS Serial Numbers: C0396
510(k) numbers
["K150873"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3003873069
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register