Hill-Rom, Inc.: Liko Sabina 200 Lift, Liko, Sweeden. The Sabina and Capella sit-to-stand lifts are especially designed for people who have difficulty in standing on their own. The lifts are intended to be used with patients who are able to bear weight on their legs and actively participate in the standing exercise.

Recall number
Z-1696-2011
Recalling firm
Hill-Rom, Inc.
Product
Liko Sabina 200 Lift, Liko, Sweeden. The Sabina and Capella sit-to-stand lifts are especially designed for people who have difficulty in standing on their own. The lifts are intended to be used with patients who are able to bear weight on their legs and actively participate in the standing exercise.
Status
Terminated
Initiated
2011-01-27
Posted
2011-03-17
Terminated
2014-09-19
Reason
The firm has received reports of injuries related to the Sabina Sit-in-Stand lifts. Of six injury reports, two were injuries in patients during use and four injuries were in non-patients when they tripped over the lift device and fell on the unit.
Root cause
Device Design
Action
The firm, Liko, issued two "URGENT FIELD SAFETY NOTICES" one dated January 27, 2011 and a follow-up notice dated February 9, 2011 to its consignees/customers. The notices described the product, problem and actions to be taken. The customers were instructed to keep the device in a separate unoccupied area when not in use; use extra care and support when lifting patients with the device; not allow unauthorized persons, especially children, to play around or operate the lift; examine the sling hooks daily to make sure they are secure, and to forward a copy of this notice to any other facility personnel they deem appropriate. The notices state the firm is investigating a design change to mitigate the risk of injuries with the device. If you have any questions concerning this request or the procedure as outlined, please contact Hill-Rom Technical Support at 800-445-3720.
Quantity
not on file
Distribution
Worldwide distribution: USA including: AL, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK ,OR, PA, RI, SC, SD, TN ,TX, UT, VA, WA, WI, and WV; and country of: Canada.
Lot, serial or model codes
All lifts from 1997 to the present.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
1824206
Firm city
Batesville
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register