Skip to the content

FDA Medical Device Recalls (openFDA)

American Contract Systems, Inc.: Various Cysto Packs

Recall number
Z-1695-2022
Recalling firm
American Contract Systems, Inc.
Product
Various Cysto Packs
Status
Open, Classified
Initiated
2022-06-16
Posted
not on file
Terminated
not on file
Reason
Product was exposed to multiple sterilization cycles without validation for multiple exposures.
Root cause
Process change control
Action
The recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022. Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm. A Field Action Response Form should be completed and returned to angela.bunn@owens-minor.com as soon as possible, even if the customer has no inventory. If you have any questions, please call 816-920-5846.
Quantity
21 units
Distribution
Distribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX
Lot, serial or model codes
Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) LSRCFHCYTURC, 2105051, 883211, 15142078 2) LSRCFHCYTURC, 2106092, 853211, 44251068 3) LSRCYSTOTURF, 2106181, 834211, 44251068 4) LSRCYSTOTURF, 2108251, 775211, 44376072 5) JECY99B, 2105182, 862211, 44108484-005 6) BHCY71H, 2107011, 819211, 44247051-005 7) SLCY61, 2108111, 777211, 15093750-004 8) BHCY31D, 2107161, 810211, 44228702-005 9) BHCY31D, 2107161, 810211, 44228704-005 10) BHCY31D, 2107161, 810211, 44228703-005 11) BHCY31D, 2109032, 763211, 44333133-005 12) BHCY31D, 2109032, 763211, 44333132-005 13) BHCY31D, 2109032, 763211, 44333131-005 14) BHCY31D, 2109032, 763211, 44333300-005 15) BHCY31D, 2109032, 763211, 44333398-005 16) BHCY31D, 2110051, 726211, 44351428-005 17) BHCY31D, 2110051, 726211, 44351427-005 18) BHCY31D, 2110051, 726211, 44351426-005 19) BHCY31D, 2110051, 726211, 44351425-005 20) UTCY45M, 2104231, 890211, 44175283-005 21) FRCY56D, 2105121, 869211, 44173978-005
510(k) numbers
not on file
PMA numbers
not on file
Product code
NWU
Firm FEI
3003419114
Firm city
Kansas City
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register