Philips Medical Systems (Cleveland) Inc: Brilliance iCT Computed Tomography X-ray system

Recall number
Z-1695-2015
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance iCT Computed Tomography X-ray system
Status
Terminated
Initiated
2015-04-13
Posted
2015-05-29
Terminated
2015-10-01
Reason
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
Root cause
Other
Action
Philips sent an Urgent - Field Safety Notice dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Healthcare is recommending to stop the use of your system until the X-ray tube mounting fasteners are replaced. In addition, a Philips Field Service Engineer (FSE) will service your system via a Mandatory Field Change Order (FCO 72800638). A Philips FSE will contact you to schedule this appointment. If you need any further information or support concerning this issue, please contact your local Philips representative or your local Philips Healthcare office. This notice has been submitted to the appropriate Regulatory Agency. Philips apologizes for any inconveniences this problem has caused your organization. For further questions please call (440) 483-7600.
Quantity
2
Distribution
Internationally to the countries of China and Norway.
Lot, serial or model codes
Model# 728306; Serial #'s: 85019 & 85023.
510(k) numbers
["K060937"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register