Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.

Recall number
Z-1694-2015
Recalling firm
Zimmer, Inc.
Product
Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.
Status
Terminated
Initiated
2015-04-29
Posted
2015-05-29
Terminated
2015-12-08
Reason
The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.
Root cause
Component design/selection
Action
Zimmer sent an Urgent Medical Device Recall letter dated April, 29, 2015, to all affected customers.On 4/29/2015. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to review the notification and ensure affected personnel are aware of the contents. Locate all affected product and quarantine them immediately. Return the recalled product along with completed Inventory Return Certification Form. Notify Zimmer of any further distribution of the affected product. In addition, identify the surgeons and hospitals that you have identified using the provided spreadsheet. For further questions call the customer call center at 1-877-946-2761.
Quantity
289,616
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Canada, Brazil, Mexico, Bahamas, Nicaragua, Singapore, China, Japan, Malaysia, Taiwan, Thailand, Korea, Spain, Belgium, Switzerland, India, Germany, England, Italy, Saudi Arabia, Lebanon, Libya, United Arab Emirates, Algeria, France, United Kingdom, Canary Islands, Iraq, Jordan, Lithuania, Morocco, Macedonia, Malta, Mauritius, Serbia, Russian Federation, and Yemen.
Lot, serial or model codes
All lots. Part numbers: 00-2253-020-45  INTERLOCKING IM SCREW 4.5MM DIA X 20MM 00-2253-022-45  INTERLOCKING IM SCREW 4.5MM DIA X 22.5MM 00-2253-025-45  INTERLOCKING IM SCREW 4.5MM DIA X 25MM 00-2253-027-45  M/DN Intramedullary Fixation 4.5mm Dia. Cortical Screw 00-2253-030-45 INTERLOCKING IM SCREW 4.5MM DIA X 30MM 00-2253-032-45  INTERLOCKING IM SCREW 4.5MM DIA X 32.5MM 00-2253-035-45 INTERLOCKING IM SCREW 4.5MM DIA X 35MM 00-2253-037-45 INTERLOCKING IM SCREW 4.5MM DIA X 37.5MM 00-2253-040-45 INTERLOCKING IM SCREW 4.5MM DIA X 40MM 00-2253-042-45  INTERLOCKING IM SCREW 4.5MM DIA X 42.5MM 00-2253-045-45  INTERLOCKING IM SCREW 4.5MM DIA X 45MM 00-2253-047-45  INTERLOCKING IM SCREW 4.5MM DIA X 47.5MM 00-2253-050-45  INTERLOCKING IM SCREW 4.5MM DIA X 50MM 00-2253-052-45  INTERLOCKING IM SCREW 4.5MM DIA X 52.5MM 00-2253-055-45 INTERLOCKING IM SCREW 4.5MM DIA X 55MM 00-2253-057-45 INTERLOCKING IM SCREW 4.5MM DIA X 57.5MM 00-2253-060-45  INTERLOCKING IM SCREW 4.5MM DIA X 60MM 00-2253-062-45 INTERLOCKING IM SCREW 4.5MM DIA X 62.5MM 00-2253-065-45 INTERLOCKING IM SCREW 4.5MM DIA X 65MM 00-2253-067-45  INTERLOCKING IM SCREW 4.5MM DIA X 67.5MM 00-2253-070-45 INTERLOCKING IM SCREW 4.5MM DIA X 70MM 00-2253-072-45  INTERLOCKING IM SCREW 4.5MM DIA X 72.5MM 00-2253-075-45  INTERLOCKING IM SCREW 4.5MM DIA X 75MM 00-2253-077-45  INTERLOCKING IM SCREW 4.5MM DIA X 77.5MM 00-2253-080-45 INTERLOCKING IM SCREW 4.5MM DIA X 80MM 00-2253-082-45 INTERLOCKING IM SCREW 4.5MM DIA X 82.5MM 00-2253-085-45  INTERLOCKING IM SCREW 4.5MM DIA X 85MM 00-2253-087-45  INTERLOCKING IM SCREW 4.5MM DIA X 87.5MM 00-2253-090-45  INTERLOCKING IM SCREW 4.5MM DIA X 90MM
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDJ
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register