Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538

Recall number
Z-1693-2008
Recalling firm
Boston Scientific Corporation
Product
iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538
Status
Terminated
Initiated
2000-04-11
Posted
2008-09-23
Terminated
2009-04-06
Reason
Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.
Root cause
Process control
Action
On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.
Quantity
120 units
Distribution
Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.
Lot, serial or model codes
UPNs: H74900036130, H749FG000940, H749ILAB120INS0, H749ILAB120INSR0, H749ILAB220INS0, H749240INS0, M004EPIL120INS0. Serial numbers 0000004082, 0000004385, 0000004386, 0000004387, 0000004388, 0000004389, 0000004390, 0000004391, 0000004392, 0000004393, 0000004396, 0000004397, 0000004398, 0000004399, 0000004416, 0000004417, 0000004418, 0000004419, 0000004420, 0000004421, 0000004422, 0000004423, 0000004424, 0000004425, 0000004426, 0000004427, 0000004428, 0000004429, 0000004430, 0000004431, 0000004432, 0000004433, 0000004434, 0000004435, 0000004436, 0000004439, 0000004441, 0000004442, 0000004443, 0000004444, 0000004445, 0000004446, 0000004447, 0000004448, 0000004453, 0000004454, 0000004455, 0000004456, 0000004468, 0000004469, 0000004472, 0000004473, 0000004480, 0000004487, 0000004488, 0000004489, 0000004490, 0000004491, 0000004494, 0000004495, 0000004496, 0000004497, 0000004498, 0000004505, 0000004506, 0000004507, 0000004512, 0000004513, 0000004514, 0000004515, 0000004516, 0000004520, 0000004521, 0000004522, 0000004523, 0000004524, 0000004525, 0000004526, 0000004529, 0000004538, 0000004539, 0000004548, 0000004549, 0000004550, 0000004551, 0000004556, 0000004557, 0000004558, 0000004559, 0000004564, 0000004565, 0000004566, 0000004567, 0000004572, 0000004573, 0000004574, 0000004575, 0000004580, 0000004581, 0000004582, 0000004583, 0000004584, 0000004585, 0000004586, 0000004587, 0000004610, 0000004702, 0000004703, 0000004704, 0000004705, 0000004717, 0000004718, 0000004719, 0000004720, 0000004726, 0000004728, 0000004729, 0000004730, 0000004731, 0000004732, 0000004737, 0000004738, 0000004751, 0000035228, 0000035229, 0000035390, 0000035391, 0000036314, 0000094718, 0009430893, 0009503458, 0009543381, 0009565345, 0009902558, 0009602559, 0009626567, 0009640091, 0009828054, 0009828055, 0009828344, 0011008866.
510(k) numbers
["K072517"]
PMA numbers
not on file
Product code
IYO
Firm FEI
2939204
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register