FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538
- Recall number
- Z-1693-2008
- Recalling firm
- Boston Scientific Corporation
- Product
- iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538
- Status
- Terminated
- Initiated
- 2000-04-11
- Posted
- 2008-09-23
- Terminated
- 2009-04-06
- Reason
- Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.
- Root cause
- Process control
- Action
- On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.
- Quantity
- 120 units
- Distribution
- Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.
- Lot, serial or model codes
- UPNs: H74900036130, H749FG000940, H749ILAB120INS0, H749ILAB120INSR0, H749ILAB220INS0, H749240INS0, M004EPIL120INS0. Serial numbers 0000004082, 0000004385, 0000004386, 0000004387, 0000004388, 0000004389, 0000004390, 0000004391, 0000004392, 0000004393, 0000004396, 0000004397, 0000004398, 0000004399, 0000004416, 0000004417, 0000004418, 0000004419, 0000004420, 0000004421, 0000004422, 0000004423, 0000004424, 0000004425, 0000004426, 0000004427, 0000004428, 0000004429, 0000004430, 0000004431, 0000004432, 0000004433, 0000004434, 0000004435, 0000004436, 0000004439, 0000004441, 0000004442, 0000004443, 0000004444, 0000004445, 0000004446, 0000004447, 0000004448, 0000004453, 0000004454, 0000004455, 0000004456, 0000004468, 0000004469, 0000004472, 0000004473, 0000004480, 0000004487, 0000004488, 0000004489, 0000004490, 0000004491, 0000004494, 0000004495, 0000004496, 0000004497, 0000004498, 0000004505, 0000004506, 0000004507, 0000004512, 0000004513, 0000004514, 0000004515, 0000004516, 0000004520, 0000004521, 0000004522, 0000004523, 0000004524, 0000004525, 0000004526, 0000004529, 0000004538, 0000004539, 0000004548, 0000004549, 0000004550, 0000004551, 0000004556, 0000004557, 0000004558, 0000004559, 0000004564, 0000004565, 0000004566, 0000004567, 0000004572, 0000004573, 0000004574, 0000004575, 0000004580, 0000004581, 0000004582, 0000004583, 0000004584, 0000004585, 0000004586, 0000004587, 0000004610, 0000004702, 0000004703, 0000004704, 0000004705, 0000004717, 0000004718, 0000004719, 0000004720, 0000004726, 0000004728, 0000004729, 0000004730, 0000004731, 0000004732, 0000004737, 0000004738, 0000004751, 0000035228, 0000035229, 0000035390, 0000035391, 0000036314, 0000094718, 0009430893, 0009503458, 0009543381, 0009565345, 0009902558, 0009602559, 0009626567, 0009640091, 0009828054, 0009828055, 0009828344, 0011008866.
- 510(k) numbers
- ["K072517"]
- PMA numbers
- not on file
- Product code
- IYO
- Firm FEI
- 2939204
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28