FDA Medical Device Recalls (openFDA)
PerkinElmer Life and Analytical Sciences, Wallac, OY: NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)06438147295708(17)180630(10)655300 (01)06438147295708(17)180630(10)654924 (01)06438147295708(17)180831(10)656025 (01)06438147295708(17)180831(10)657055 (01)06438147295708(17)180831(10)656170 (01)06438147295708(17)180831(10)656320 (01)06438147295708(17)181031(10)657768 (01)06438147295708(17)181031(10)657513 (01)06438147295708(17)181031(10)658006
- Recall number
- Z-1692-2018
- Recalling firm
- PerkinElmer Life and Analytical Sciences, Wallac, OY
- Product
- NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)06438147295708(17)180630(10)655300 (01)06438147295708(17)180630(10)654924 (01)06438147295708(17)180831(10)656025 (01)06438147295708(17)180831(10)657055 (01)06438147295708(17)180831(10)656170 (01)06438147295708(17)180831(10)656320 (01)06438147295708(17)181031(10)657768 (01)06438147295708(17)181031(10)657513 (01)06438147295708(17)181031(10)658006
- Status
- Terminated
- Initiated
- 2017-11-02
- Posted
- not on file
- Terminated
- 2019-09-17
- Reason
- There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.
- Root cause
- Process control
- Action
- Actions to be taken: - Inspect your inventory of the NeoBase Succinylacetone Assay Solution vials for the affected lots. - If you notice leakage or deformed stopper seals, destroy the affected vials according to local instructions and regulations. - Complete the Response Form with the quantity and identity of affected Succinylacetone Assay Solution vials in your inventory. - Return the Response Form to PerkinElmer. - PerkinElmer will replace the quantity of leaking Succinylacetone Assay Solution vials found and destroyed from your inventory
- Quantity
- 5275
- Distribution
- Distributed in 20 US states: AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MI, MO, NV, OH, OR, SC, TN, TX.
- Lot, serial or model codes
- Affected lots: 654184 653414 653657 654923 655680 655300 654924 656025 657055 656170 656320 657768 657513 658006
- 510(k) numbers
- ["K083130"]
- PMA numbers
- not on file
- Product code
- NQL
- Firm FEI
- 3002808498
- Firm city
- Turku
- Firm state
- not on file
- Firm country
- Finland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28