FDA Medical Device Recalls (openFDA)

Smiths Medical ASD, Inc.: Smiths Medical, BCI Advisor Vital Signs Monitor, Product Reorder No 92M774325A.

Recall number
Z-1692-2015
Recalling firm
Smiths Medical ASD, Inc.
Product
Smiths Medical, BCI Advisor Vital Signs Monitor, Product Reorder No 92M774325A.
Status
Terminated
Initiated
2015-03-21
Posted
2015-05-29
Terminated
2015-11-20
Reason
The Auxiliary Serial Input/ Output Modes on the Advisor do not function as expected, resulting in data formatting issues or incorrect data being transmitted in or out of the serial port into a data collection system.
Root cause
Nonconforming Material/Component
Action
Consignee was sent on 3/20/2015 a Smiths Medical "Urgent Medical Device Recall" letter dated 13 March 2015. The letter was addressed to Supplier Quality. The letter described the issue and the product affected by the recall. Requested consignees to quarantine the affected product and to complete and return the Confirmation Form by Fax to 1-800-237-8033 or by email to recall.response@smiths-medical.com within 10 days of receipt of this letter. Customers shall report any issues with these products to Smiths Medical's Global Complaint Department at 1-866-216-8806 or globalcompaints@smiths-medical.com. For questions they can contact Smiths Medical's Customer Service Department at 1-800-258-5361. UPDATE 4-20-2015 Smiths Medical sent an Urgent Medical Device Recall letter via email to the one consignee on April 10, 2015. The letter identified the additional monitors and requested that those devices be quarantined, return the attached confirmation form, and a customer service representative will contact the consignee to arrange for exchange of the affected Advisor monitors for credit or replacement.
Quantity
5 UPDATE 4-20-2015 to include 10 additional devices
Distribution
MN
Lot, serial or model codes
Serial Numbers: 4059587 through 4059591  UPDATED 4-20-2015: additional devices affected have serial numbers: 4059949, 4059950, 4059951, 4059952, 4059953, 4059954, 4059955, 4059957, 4059958 and 4059960
510(k) numbers
["K030407"]
PMA numbers
not on file
Product code
CCK
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register