Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Prolieve Thermodilatation Catheter Kit, a Transurethral Microwave Therapy (TUMT) System, for BPH (Benign Prostatic Hyperplasia) Thermotherapy, Material Numbers M0068808022 and M0068808170.

Recall number
Z-1692-2009
Recalling firm
Boston Scientific Corporation
Product
Prolieve Thermodilatation Catheter Kit, a Transurethral Microwave Therapy (TUMT) System, for BPH (Benign Prostatic Hyperplasia) Thermotherapy, Material Numbers M0068808022 and M0068808170.
Status
Terminated
Initiated
2009-07-20
Posted
2009-07-28
Terminated
2010-03-04
Reason
Balloon leaks: Boston Scientific has received complaints from the field involving product leaks associated with the anchor balloons and compression balloons. Correspondence between BSC and FDA has led to BSC's decision to remove the Prolieve Catheter Kits from the field.
Root cause
Process design
Action
Boston Scientific sent Customer notification letters via Federal Express Priority mail on July 20, 2009. A total of 812 US consignees were notified and a total of 26 consignees in Puerto Rico. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product in their inventory. Follow-up communications will be made to the accounts that do not respond to this inquiry.
Quantity
26,935 kits
Distribution
Nationwide distribution, including Puerto Rico.
Lot, serial or model codes
Material Number M0068808022 with Lot/Batch Numbers: 604951, 604952, 604953, 604954, 604955, 604956, 604957, 604958, 604959, 604960, 605015, 605016, 605017, 605276, 605277, 605278, 605279, 605280, 605281, 605282, 605283, 605284, 605285, 605286, 605287, 605288, 605289, 605290, 605291, 605292, 605293, 605294, 605295, 605296, 605297, 605298, 605299, 605783, 605784, 605785, 605786, 605787, 605788, 605789, 605790, 605791, 606044, 606045, 606046, 606079, 606080, 606081, 606082, 606083, 606084, 606152, 606182, 606183, 606203, 606299, 606300, 606301, 606302, 606303, 606304, 606420, 606421, 606422, 606423, 606424, 606597, 606598, 606599, 606600, 606601, 606602, 606603, 606604, 606605, 606606, 606607, 606781, 606782, 606829, 606830, 606831, 606832, 606833, 606834, 606879, 606880, 607014, 607015, 607016, 607017, 607018, 607019, 607020, 607021, 607022, 607023, 607024, 607025, 607189, 607190, 607191, 607192, 607193, 607194, 607315, 607316, 607317, 607318, 607321, 607322, 607323, 607324, 607325, 607520, 607521, 607522, 607523, 607524, 607525, 607526, 607527, 607528, 607646, 607735, 607736, 607737, 607738, 607739, 607740, 608160, 608161, 608162, 608163, 608164, 608165, 608166, 608167, 608168, 608169, 608575, 608591, 608592, 608593, 608594, 608595, 608596, 608597, 608602, 608680, 608681, 619219, 619221, 619222, 619311, 619312, 619313, 623435, 623436, 627518, 627519, 628502, 628503, 628504, 633348, 633349, 633350, 637211, 637212, 637216, 637217, 637220, 639295, 639296, 642203, 642204, 642549, 642576, 644535, 644536, 644537, 647635, 647636, 650229, 650230, 650232, 650233, 650234, 651870, 651871, 651872, 653449, and M19406. Material Number M0068808170 with Lot/Batch Numbers: 608608, 608609, 608610, 608611, 608612, 651734, 651874, and 651877.
510(k) numbers
not on file
PMA numbers
["P030006"]
Product code
MEQ
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28