Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT

Recall number
Z-1691-2026
Recalling firm
Medline Industries, LP
Product
Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
Status
Open, Classified
Initiated
2026-02-13
Posted
2026-03-27
Terminated
not on file
Reason
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Root cause
Process change control
Action
On February 13, 2026, the firm notified affected consignees via Urgent Medical Device Recall letters delivered through First Class mail and email. Customers were instructed to destroy all affected ENFit G-Tube connectors and kits that contain the G-Tube connectors.
Quantity
9,040 kits
Distribution
Worldwide - US Nationwide and the countries of Canada.
Lot, serial or model codes
1. UDI-DI 10198459016257 (ea) 40198459016258 (case) 2. UDI-DI 10193489860078 (ea) 40193489860079 (case) ALL LOTS
510(k) numbers
not on file
PMA numbers
not on file
Product code
PIO
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register