FDA Medical Device Recalls (openFDA)
Davol, Inc., Subs. C. R. Bard, Inc.: Bard Ventralight" ST Mesh with Echo PS" 4"x6" Reorder Number: 5955460 Indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.
- Recall number
- Z-1690-2014
- Recalling firm
- Davol, Inc., Subs. C. R. Bard, Inc.
- Product
- Bard Ventralight" ST Mesh with Echo PS" 4"x6" Reorder Number: 5955460 Indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.
- Status
- Terminated
- Initiated
- 2014-04-24
- Posted
- 2014-06-02
- Terminated
- 2015-12-23
- Reason
- Pouch holding sterile inflation assembly and inflation adapter may be open or have a weak seal.
- Root cause
- Packaging process control
- Action
- Davol, Inc., Subs. C. R. Bard, Inc.sent an Urgent Medical Device Recall letter dated April 28, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to examine their inventory and quarantine product subject to the recall, contact Davol Customer Service at 1-800-556-6275 or C.R. Bard's Medical Services and Support Department at 1-800-562-0027 for instructions on how to return the product to Davol, Consignees were also instructed to complete and return the recall effectiveness check form, and notify their customers if the product was further distributed. For questions regarding this recall call 1-800-556-6275.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to AUSTRIA, BELGIUM, CANADA, CYPRUS, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ITALY, LUXEMBOURG, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, and UNITED KINGDOM
- Lot, serial or model codes
- Lot Number: HUXH0695
- 510(k) numbers
- ["K122436"]
- PMA numbers
- not on file
- Product code
- GCJ
- Firm FEI
- 1213643
- Firm city
- Warwick
- Firm state
- RI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28