Codman & Shurtleff, Inc.: MedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine Product Code:91-4200 (20ml pump) US 91-4200. Intended for the intrathecal delivery of Baclofen or Morphine.
- Recall number
- Z-1690-2013
- Recalling firm
- Codman & Shurtleff, Inc.
- Product
- MedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine Product Code:91-4200 (20ml pump) US 91-4200. Intended for the intrathecal delivery of Baclofen or Morphine.
- Status
- Terminated
- Initiated
- 2013-06-07
- Posted
- 2013-07-16
- Terminated
- 2015-04-21
- Reason
- Miscalibrated Fill Level Sensor may affect dosing level
- Root cause
- Device Design
- Action
- Codman issued a Medical Device Correction Notification to Medical Facilities dated June 7, 2013 via FedX to inform them of the potential problem and a Worksheet (instructions) to Identify Pumps with Miscalibrated FLS during each patient's next scheduled refill session or sooner if patient is symptomatic. If a pump is identified as having a miscalibrated FLS, notify Codman Neuro using the same worksheet. In addition a sample letter was provided to Physi cans to provide to patients on the system. Follow the recommendations for managing patients who are identified as having a pump with a miscalibrated FLS. For additional information,contact your Codman Neuro representative, or contact Codman Neuro Clinical Support at 1-800-660-2660. Please report any malfunction or adverse event related to this device to Codman Neuro at 1-866-491-0974, Option 2.
- Quantity
- 2 units US and 735 OUS
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of FL, NC, NJ, and TX, and the countries of Austria, Belgium, Spain, Great Britain, Germany, Italy, Ireland, Luxemburg, Russia, Saudi Arabia, Portugal, Greece, Egypt, Switzerland, and Turkey.
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- not on file
- PMA numbers
- ["P890055"]
- Product code
- LKK
- Firm FEI
- 3002947440
- Firm city
- Raynham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28