Stryker Howmedica Osteonics Corp.: Universal Notch Preparation Guide, Size #3; Non Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430; authorized representative in Europe: Stryker France ZAC Satolas Green Pusignan; Cedex, France.

Recall number
Z-1690-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Universal Notch Preparation Guide, Size #3; Non Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430; authorized representative in Europe: Stryker France ZAC Satolas Green Pusignan; Cedex, France.
Status
Terminated
Initiated
2008-03-18
Posted
2008-09-20
Terminated
2010-10-25
Reason
Discovery of the potential for Pin and Punch interference of the Size 3 Scorpio Notch Preparation Guide that may result in damage to bone if a user aggressively attempts to seat an obstructed punch.
Root cause
Device Design
Action
Urgent Product Correction notification letters were sent via Federal Express, return receipt to all users on March 18, 2008. Questions should be directed to Rita Intorella, Divisional Regulatory Reporting at (201) 831-5825.
Quantity
739 units in US; 1213 units internationally
Distribution
Products were distributed worldwide.
Lot, serial or model codes
All lots manufactured since 1998. Catalog Number 7650-3363. Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430; authorized representative in Europe: Stryker France ZAC Satolas Green Pusignan; Cedex, France.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register