FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: ADVIA Centaur DHEA-SO4
- Recall number
- Z-1688-2018
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- ADVIA Centaur DHEA-SO4
- Status
- Terminated
- Initiated
- 2018-01-02
- Posted
- 2018-05-04
- Terminated
- 2023-02-10
- Reason
- The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
- Root cause
- Error in labeling
- Action
- Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
- Quantity
- not on file
- Distribution
- Nationwide and Canada, Mexico, and Thailand
- Lot, serial or model codes
- ADVIA Centaur¿ DHEA-SO4; radioimmunoassay, dehydroepiandrosterone (free and sulfate); SMN 10282227 ADVIA Centaur DHEA-S Kit Lot # 00119053 14041053 80449053 90520053 29750054 30423054 39155054 50533054 63267055 68936055 79273055 86427055 06402057 07109057 14187057 37499057
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JKC
- Firm FEI
- 2432235
- Firm city
- Tarrytown
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28