James Leckey Design Ltd: MyWay Walker Chassis - Product Usage: is intended for use by children already using a supportive walking device.
- Recall number
- Z-1687-2021
- Recalling firm
- James Leckey Design Ltd
- Product
- MyWay Walker Chassis - Product Usage: is intended for use by children already using a supportive walking device.
- Status
- Open, Classified
- Initiated
- 2021-03-01
- Posted
- not on file
- Terminated
- not on file
- Reason
- Due to a failure of the welds, the casters detached from the rear legs of the walker.
- Root cause
- Device Design
- Action
- On 03/29/2021, the Recalling Firm and its U.S. Distributor sent an "URGENT: MEDICAL DEVICE RECALL" Notification Letter via mail informing customers that: 1. The rear legs of the walker each have a boss into which castors are screwed. The Recalling Firm has had reports of these bosses coming loose and castors detaching from the rear legs. 2. The rear leg assembly consists of an aluminum tube with bosses welded at either end. Our investigations have identified a common cause of the field issues as a failure of the weld between the tube and the bottom boss. 3. The weld is fabricated by welding through a hole in the tube to create the bond and the weld is then continued to completely fill the hole. The boss is welded at either side of the tube, and the welds are then smoothed off on the outside of the tube prior to painting, obscuring the weld on the finished product. The failure is due to variations in the manual weld process, primarily due to variability in the weld parameters used. This led to the solid aluminum boss not being heated sufficiently in the weld process, thereby creating a weak bond. Process controls did not prevent the process failure or detect that there was an issue prior to dispatch. 4. On evaluation of the failure, it is seen if the weld breaks, initially the castor boss will be retained in the tube by the roughened weld surface between the tube and boss. However over, the boss will potentially work its way loose, or through repeated use the roughening will be worn away and allow the castor boss to come loose. Use of this product may result in the unexpected detachment of the caster which could result in a fall and potential injury due to a fall. Customers are instructed to: -Immediately examine their inventory and quarantine product subject to the recall. -If they further distributed the affected products, notify them of the product recall by forwarding/including a copy of this recall notification. -Complete and return the Acknowle
- Quantity
- 2,719 walkers
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of CA and WI and the countries of Italy, Norway, Australia, Czechia, Netherlands, Denmark, New Zealand, Japan, United Arab Emirates, Qatar, Croatia, Spain, Poland, Russia, South Africa, Taiwan, Saudi Arabia, Chemitec, Centre Harika, Hong Kong, Lebanon, Belgium, Philippines, Singapore, Malaysia, Indonesia, Slovakia, China, and France.
- Lot, serial or model codes
- Catalog Codes/Item Number: S1 - 173-1600/PMWS1x; S2 - 173-2600/PMWS2x ; S3 - 173-3600/PMWS3x; Serial Number Range: 220568 to 272111 manufactured between 08/01/2018 to 11/26/2020
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- ITJ
- Firm FEI
- 3010712784
- Firm city
- Lisburn
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28