Abbott Vascular: Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Recall number
Z-1687-2019
Recalling firm
Abbott Vascular
Product
Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
Status
Terminated
Initiated
2019-05-01
Posted
not on file
Terminated
2020-12-11
Reason
Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.
Root cause
Use error
Action
On 05/01/19, Urgent Medical Device Correction notices were sent to implanting physicians, via personal delivery or email, informing them that under normal use conditions, this failure will not occur. To prevent unintended force from being applied to the clip, revised instructions, including definition and related technique, for performing the steps "Establish Final Angle" and "Invert the Clip Arms" were developed and provided herein. The revised instructions will be applied to the shared XTR and NTR Clip Delivery System Instructions for Use. Unintended excessive force applied to the NTR Clip Delivery System during "Establish Final Arm Angle" can result in unexpected movement of clip arms; however, unlike XTR, the NTR Clip does not become damaged and remains functional. Adhering to the revised IFU sections for both product platforms will ensure consistency of use. Further, the firm requested that customers read the Field Safety Notice, review revised instructions for use with firm representatives, sign and return Effectiveness Check Forms, and share this information with all personnel associated with the procedure. The revised instruction steps provided in this communication will be updated in the product IFU and included in associated training. Your current inventory of product is acceptable for safe use following the revised IFU steps described above. There is no need to return any product. Customers with additional questions are encouraged to call customer service at 800-227-9902.
Quantity
not on file
Distribution
U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM
Lot, serial or model codes
All Serial Numbers.
510(k) numbers
not on file
PMA numbers
["P100009S025"]
Product code
NKM
Firm FEI
2024168
Firm city
Temecula
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register