Diamedix Corporation: Diamedix Is-EBV-EA-D IgG Test Kit Diamedix Is-Mumps IgG Test Kit is an EIA procedure intended for the qualitative and semi-quantitative detection of Mumps IgG antibodies as an aid in the diagnosis of Mumps infection.
- Recall number
- Z-1687-2016
- Recalling firm
- Diamedix Corporation
- Product
- Diamedix Is-EBV-EA-D IgG Test Kit Diamedix Is-Mumps IgG Test Kit is an EIA procedure intended for the qualitative and semi-quantitative detection of Mumps IgG antibodies as an aid in the diagnosis of Mumps infection.
- Status
- Terminated
- Initiated
- 2014-10-20
- Posted
- not on file
- Terminated
- 2016-12-14
- Reason
- A gel-like substance was found in kit component, sample diluent , that could potentially compromise product performance.
- Root cause
- Unknown/Undetermined by firm
- Action
- The firm, Diamedix, sent letters dated October 14, October 20 and December 5, 2014 to its customers directly informing them of the correction action. Notification of all domestic customers via Certified Mail. Notification fo the international customers via Federal Express and/or e-mail. With the letter, a replacement bottle of Sample B Diluent was sent and the customer was advised to discard any unused bottles of affected product. New lot number labels were provided to identify kits with the replacement Sample B Diluent. Customers with questions were told they could contact Diamedix Technical Services (1-800-327-4565) if they had any questions or required any additional guidance regarding the correction.
- Quantity
- 94 kits
- Distribution
- Worldwide distribution: US nationwide including states of: AZ, CA, CO, FL, GA, IL, IN, KY, MA, MI, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI and Puerto Rico. and Country of: ITALY.
- Lot, serial or model codes
- Lot # 51104, Catalog # 720-640 Expiration Date August 31, 2015
- 510(k) numbers
- ["K981831"]
- PMA numbers
- not on file
- Product code
- GNP
- Firm FEI
- 1044713
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28