Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corporation: MEGA 8Fr 50cc Intra-Aortic Balloon Catheter

Recall number
Z-1685-2022
Recalling firm
Datascope Corporation
Product
MEGA 8Fr 50cc Intra-Aortic Balloon Catheter
Status
Open, Classified
Initiated
2022-08-05
Posted
2022-09-08
Terminated
not on file
Reason
Certain lots containing undersized dilator.
Root cause
Vendor change control
Action
The firm distributed an Urgent Medical Device Removal notification, by mail, dated August 5, 2022. The notification informs customers that certain lots of 8Fr. 50cc IAB catheter and insertion kits are being recalled due to an undersized vessel dilator. During a sheathless insertion, the undersized inner diameter may result in the inability to insert the guidewire through the inner lumen of the dilator and the undersized outer diameter may reduce dilation of the vessel. If used in a procedure, recalled product could result in a procedural delay, gas emoli, and vessel injury. Customers are asked to examine their inventory to determine if they have nay recalled product and to quarantine any affected units. Customers with recalled product are to contact Datascope Corp./Getinge Customer Service at 888-9GETUSA (888-943-8872, option 2) between the hours of 8 AM and 6 PM to request a return authorization number to receive full credit. Distributors who have shipped any product are to forward the notification to customers. The recall notification is to be forwarded to all users at hospitals/facilities. Customers are asked to complete and return the recall response form by email to Dilator8Fr2022@getinge.com or by fax to 877-647-1921.
Quantity
10,427 units
Distribution
Awaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Iceland, Ireland, Italy, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Lot, serial or model codes
Code information reads "UDI-DI, Lot No., Finished Goods Part No.": 10607567109619, 3000184437, 0684-00-0296-01U; 10607567107301, 3000171501, 0684-00-0498-01; 10607567107301, 3000183312, 0684-00-0498-01; 10607567107301, 3000183313, 0684-00-0498-01; 10607567107301, 3000186654, 0684-00-0498-01; 10607567107301, 3000186655, 0684-00-0498-01.
510(k) numbers
["K133074"]
PMA numbers
not on file
Product code
DSP
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register