Teleflex Medical: RUSCH LASERTUBE (Rubber), Order No. : 102004-000070, REF 102004, SIZE I.D. mm 7,0

Recall number
Z-1685-2019
Recalling firm
Teleflex Medical
Product
RUSCH LASERTUBE (Rubber), Order No. : 102004-000070, REF 102004, SIZE I.D. mm 7,0
Status
Terminated
Initiated
2019-04-01
Posted
not on file
Terminated
2025-06-03
Reason
The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges.
Root cause
Process control
Action
The firm initiated the recall by letter on 04/01/2019. The letter stated reason for the recall and requested the return of the recalled units. Distributors were directed to collect the recalled units from their customers, consolidate them, and complete and return the acknowledgement form. The recalled units will be returned.
Quantity
4312 units
Distribution
Nationwide
Lot, serial or model codes
Lot Numbers: 14161, 14171, 14201, 14211, 14231, 14261, 14301, 14311, 14361, 14431, 14481, 14491, 15041, 15051, 15081, 15091, 15121, 15131, 15161, 15171, 15201, 15221, 15241, 15261, 15371, 15411, 15421, 15501, 16031, 16071, 16111, 16151, 16161, 16181, 16201, 16221, 16241, 16251, 16271, 16281, 16291, 16301, 16351, 16361, 16371, 16381, 16391, 16401, 16421, 16441, 16461, 16471, 16481, 17061, 17071, 17161, 17191, 17211, 17221, 17251, 17271, 17291, 17301, 17311, 17381, 17391, 17411, 17471, 17481, 17511, 18021, 18081, 18091, 18151, 18161, 18211, 18241, 18291, 18381, 18411, 18421, 18431, 18441
510(k) numbers
["K935842"]
PMA numbers
not on file
Product code
BTR
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register