Skip to the content

FDA Medical Device Recalls (openFDA)

Diamedix Corporation: Diamedix Is-Mumps IgG Test Kit Diamedix Is-Mumps IgG Test Kit is an EIA procedure intended for the qualitative and semi-quantitative detection of Mumps IgG antibodies as an aid in the diagnosis of Mumps infection.

Recall number
Z-1685-2016
Recalling firm
Diamedix Corporation
Product
Diamedix Is-Mumps IgG Test Kit Diamedix Is-Mumps IgG Test Kit is an EIA procedure intended for the qualitative and semi-quantitative detection of Mumps IgG antibodies as an aid in the diagnosis of Mumps infection.
Status
Terminated
Initiated
2014-10-20
Posted
not on file
Terminated
2016-12-14
Reason
A gel-like substance was found in kit component, sample diluent , that could potentially compromise product performance.
Root cause
Unknown/Undetermined by firm
Action
The firm, Diamedix, sent letters dated October 14, October 20 and December 5, 2014 to its customers directly informing them of the correction action. Notification of all domestic customers via Certified Mail. Notification fo the international customers via Federal Express and/or e-mail. With the letter, a replacement bottle of Sample B Diluent was sent and the customer was advised to discard any unused bottles of affected product. New lot number labels were provided to identify kits with the replacement Sample B Diluent. Customers with questions were told they could contact Diamedix Technical Services (1-800-327-4565) if they had any questions or required any additional guidance regarding the correction.
Quantity
697 kits
Distribution
Worldwide distribution: US nationwide including states of: AZ, CA, CO, FL, GA, IL, IN, KY, MA, MI, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI and Puerto Rico. and Country of: ITALY.
Lot, serial or model codes
Lot # 20804, Catalog # 720-540. Expiration Date August 31, 2015
510(k) numbers
["K991274"]
PMA numbers
not on file
Product code
LJY
Firm FEI
1044713
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register