FDA Medical Device Recalls (openFDA)
Diamedix Corporation: Is-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease.
- Recall number
- Z-1684-2008
- Recalling firm
- Diamedix Corporation
- Product
- Is-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease.
- Status
- Terminated
- Initiated
- 2008-03-21
- Posted
- 2008-09-16
- Terminated
- 2009-02-11
- Reason
- Standards and controls in the Is-anti-Gliadin IgG Test Kits are exhibiting low activity. A drop in the O.D.s the Positive Control could potentially recover low out of range (which would invalidate the test), or false positive test results could be obtained.
- Root cause
- Environmental control
- Action
- A Customer Notification Letter was sent to each customer via Certified Mail for domestic customers and via Federal Express for international customer requesting them to contact Diamedix Technical Services to arrange for the return and replacement of any kits of the lot in question. If you have any questions please contact Glenn Gerstenfeld at 1-800-327-4565 or (305) 324-2300.
- Quantity
- 174 Test Kits
- Distribution
- Class II Recall - Worldwide Distribution.
- Lot, serial or model codes
- Catalog Number # 720-800 -- Lot # S0207, Expiration Date: Nov. 30, 2008.
- 510(k) numbers
- ["K012795"]
- PMA numbers
- not on file
- Product code
- MST
- Firm FEI
- 1044713
- Firm city
- Miami
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28