Skip to the content

FDA Medical Device Recalls (openFDA)

Diamedix Corporation: Is-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease.

Recall number
Z-1684-2008
Recalling firm
Diamedix Corporation
Product
Is-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease.
Status
Terminated
Initiated
2008-03-21
Posted
2008-09-16
Terminated
2009-02-11
Reason
Standards and controls in the Is-anti-Gliadin IgG Test Kits are exhibiting low activity. A drop in the O.D.s the Positive Control could potentially recover low out of range (which would invalidate the test), or false positive test results could be obtained.
Root cause
Environmental control
Action
A Customer Notification Letter was sent to each customer via Certified Mail for domestic customers and via Federal Express for international customer requesting them to contact Diamedix Technical Services to arrange for the return and replacement of any kits of the lot in question. If you have any questions please contact Glenn Gerstenfeld at 1-800-327-4565 or (305) 324-2300.
Quantity
174 Test Kits
Distribution
Class II Recall - Worldwide Distribution.
Lot, serial or model codes
Catalog Number # 720-800 -- Lot # S0207, Expiration Date: Nov. 30, 2008.
510(k) numbers
["K012795"]
PMA numbers
not on file
Product code
MST
Firm FEI
1044713
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register