Qiagen Sciences LLC: QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223
- Recall number
- Z-1683-2023
- Recalling firm
- Qiagen Sciences LLC
- Product
- QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223
- Status
- Open, Classified
- Initiated
- 2023-04-28
- Posted
- 2023-06-02
- Terminated
- not on file
- Reason
- Faulty cartridges in the lot could result in false test results.
- Root cause
- Under Investigation by firm
- Action
- An Urgent Medical Device Correction notification letter dated April 2023 was sent to customers. Actions to be taken by customer/user Please check if you have remaining stock of LOT 175011350. Please DO NOT use it. Dispose of it immediately in accordance to your national and local safety and environmental regulations. Please contact QIAGEN Technical Services for a free-of-charge replacement. If you already used cartridges from this LOT, please identify the results obtained with cartridges from the affected SN. For results obtained with an affected SN, review the results as follows: o Review all results of the respective targets mentioned in this notice to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. o Review any result for which tests were performed on clinical and/or epidemiological suspicion of the respective targets. Review this notice with your laboratory/medical director. IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. Complete the Acknowledgement of Receipt Form attached to this letter by 14 May 2023. Actions taken by QIAGEN QIAGEN was able to identify the root cause of the issue and has implemented immediate actions to ensure that the currently produced cartridges will not cause this issue moving forward. If you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
- Quantity
- 347 kits
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, MD, MS, NY, OK, PA, TN.
- Lot, serial or model codes
- GTIN 14053228038846; Lot 175011350; Serial Numbers within the range of 330568064 to SN 330568597 AND 330569655, 330569651, 330569650, 330569646, 330569645, 330569644, 330569643, 330569641.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- QLT
- Firm FEI
- 3003572099
- Firm city
- Germantown
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28