Skip to the content

FDA Medical Device Recalls (openFDA)

Alere Scarborough, Inc. dba Binax, Inc.: Alere i Influenza A & B Product Usage: The Alere i Influenza A and B assay, performed on the Alere i Instrument, is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans, in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.

Recall number
Z-1677-2015
Recalling firm
Alere Scarborough, Inc. dba Binax, Inc.
Product
Alere i Influenza A & B Product Usage: The Alere i Influenza A and B assay, performed on the Alere i Instrument, is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans, in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.
Status
Terminated
Initiated
2015-04-22
Posted
2015-05-26
Terminated
2016-01-14
Reason
High invalid rate and an increased risk of false negative results with certain lots of the Alere i Influenza A & B kit.
Root cause
Under Investigation by firm
Action
Alere sent an Urgent Medical Device Recall letter dated April 23, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to stop using the product and discard any unused product. Customers asked to complete and return the Verification form within 10 days to confirm receipt of the recall notification and to request credit. For questions about the information contained in the notification, contact Alere Technical Services, Monday through Friday, 8am to 8pm EST. Telephone: (855) 731-2288 or E-mail: t s.scr@alere.com. For questions regarding the status of your credit, please contact your AI ere i distributor or E-mail: ssc.alerefieldactions@alere.com.
Quantity
213 kits plus 3 - evaluation use only kits
Distribution
Worldwide Distribution - US Nationwide in the states of AL CA CT FL IA ID IL IN KY LA MD MO MS NC NE NJ OH PA SC TN TX and the countries of: Hong Kong and Singapore.
Lot, serial or model codes
PN 425-024, LN 0072633 (United States Distribution) and PN 425-000, LN 0073888 (distributed in Hong Kong, Singapore - evaluation use only)
510(k) numbers
["K141520"]
PMA numbers
not on file
Product code
OCC
Firm FEI
1221359
Firm city
Scarborough
Firm state
ME
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register