Siemens Medical Solutions USA, Inc.: Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.

Recall number
Z-1677-2009
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
Status
Terminated
Initiated
2009-04-22
Posted
2009-08-04
Terminated
2010-06-21
Reason
The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan.
Root cause
Finished device change control
Action
Siemens sent Urgent Field Correction Recall letters dated 4/22/09 to the affected users of the Symbia S and Symbia T Systems. The accounts were informed of the potential faulty pinion gear which could crack and cause the gantry to spin freely, damaging the system and potentially injuring a patient during a scan. The customers were advised to stop using the system if they hear any abnormal grinding noise from the gantry. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a replacement of the pinion gear free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance. The accounts were requested to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter.
Quantity
38 units
Distribution
Worldwide Distribution -- USA, Australia, Austria, Brazil, Bulgaria, Canada, China, Finland, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Russia, Spain, Taiwan, and Turkey.
Lot, serial or model codes
part 10275007, serial numbers 1043, 1044, 1048, 1049, 1050, 1051, 1052, 1053, 1054, 1055, 1056, 1057, 1058;  part 10275008, serial numbers 1044, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1063, 1064, 1065;  part 10275009, serial numbers 1037, 1038, 1039, 1040, 1041, 1042, 1044, 1045;  part 10275010, serial numbers 1009, 1010, 1011, and 1012.
510(k) numbers
["K041166"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register