BioReference Health, LLC: The 4Kscore Test is an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following information: a patients age, previous biopsy, and digital rectal exam (DRE). Patients file (sample type, prior biopsy status, DRE status and age) are uploaded to the Specimen Processing Module (SPM). The results from the four assays are exported automatically to Lab Manager Application (B2 LIS) which automatically triggers the 4Kscore calculation when values of the four analytes are available. The 4Kscore test results are determined automatically by the validated laboratory information system algorithm calculation software.
- Recall number
- Z-1676-2023
- Recalling firm
- BioReference Health, LLC
- Product
- The 4Kscore Test is an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following information: a patients age, previous biopsy, and digital rectal exam (DRE). Patients file (sample type, prior biopsy status, DRE status and age) are uploaded to the Specimen Processing Module (SPM). The results from the four assays are exported automatically to Lab Manager Application (B2 LIS) which automatically triggers the 4Kscore calculation when values of the four analytes are available. The 4Kscore test results are determined automatically by the validated laboratory information system algorithm calculation software.
- Status
- Open, Classified
- Initiated
- 2022-11-18
- Posted
- 2023-06-02
- Terminated
- not on file
- Reason
- SPM software anomalies that may lead to the generation of erroneous 4Kscore Test results
- Root cause
- Software Design Change
- Action
- Customers were notified via telephone between 11/18/22 and 12/31/22. In addition, the following information was provided within the corrected test results sent to customers: The 4Kscore result has been corrected to reflect the history of a prostatic nodule and/or the prior negative biopsy result, which was inaccurately posted into the algorithm, due to a bioinformatics error.
- Quantity
- 1 system (662 Patients test results affected)
- Distribution
- US Nationwide distribution in the state of New Jersey.
- Lot, serial or model codes
- UDI: 00850038860004
- 510(k) numbers
- not on file
- PMA numbers
- ["P190022"]
- Product code
- QRF
- Firm FEI
- 3003652672
- Firm city
- Elmwood Park
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28