Skip to the content

FDA Medical Device Recalls (openFDA)

BioReference Health, LLC: The 4Kscore Test is an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following information: a patients age, previous biopsy, and digital rectal exam (DRE). Patients file (sample type, prior biopsy status, DRE status and age) are uploaded to the Specimen Processing Module (SPM). The results from the four assays are exported automatically to Lab Manager Application (B2 LIS) which automatically triggers the 4Kscore calculation when values of the four analytes are available. The 4Kscore test results are determined automatically by the validated laboratory information system algorithm calculation software.

Recall number
Z-1676-2023
Recalling firm
BioReference Health, LLC
Product
The 4Kscore Test is an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following information: a patients age, previous biopsy, and digital rectal exam (DRE). Patients file (sample type, prior biopsy status, DRE status and age) are uploaded to the Specimen Processing Module (SPM). The results from the four assays are exported automatically to Lab Manager Application (B2 LIS) which automatically triggers the 4Kscore calculation when values of the four analytes are available. The 4Kscore test results are determined automatically by the validated laboratory information system algorithm calculation software.
Status
Open, Classified
Initiated
2022-11-18
Posted
2023-06-02
Terminated
not on file
Reason
SPM software anomalies that may lead to the generation of erroneous 4Kscore Test results
Root cause
Software Design Change
Action
Customers were notified via telephone between 11/18/22 and 12/31/22. In addition, the following information was provided within the corrected test results sent to customers: The 4Kscore result has been corrected to reflect the history of a prostatic nodule and/or the prior negative biopsy result, which was inaccurately posted into the algorithm, due to a bioinformatics error.
Quantity
1 system (662 Patients test results affected)
Distribution
US Nationwide distribution in the state of New Jersey.
Lot, serial or model codes
UDI: 00850038860004
510(k) numbers
not on file
PMA numbers
["P190022"]
Product code
QRF
Firm FEI
3003652672
Firm city
Elmwood Park
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register