FDA Medical Device Recalls (openFDA)
Oculus Optikgeraete GMBH: Pentacam HR REF 70900. Used to image the anterior segment of the eye.
- Recall number
- Z-1674-2022
- Recalling firm
- Oculus Optikgeraete GMBH
- Product
- Pentacam HR REF 70900. Used to image the anterior segment of the eye.
- Status
- Open, Classified
- Initiated
- 2022-07-08
- Posted
- not on file
- Terminated
- not on file
- Reason
- Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs
- Root cause
- Software design
- Action
- On 07/08/2022, OCULUS GmbH issued an "Urgent safety information Corrective Action " Letter to its subsidiary OCULUS Inc via UPS express which identifies a software error in the software version 1.26r26 and 1.26r27 in the IOL calculator of the Pentacam software. Corrective measures to be taken: A software update to version 1.26r28 must be performed on all computers used and affected in connection with the Pentacam and additional workstations. Users should refrain from using the printout of the IOL calculator for affected installations until after the software update has been carried out. For questions or support - contact OCULUS Inc Service Team at 425-670-9977 or email service@oculususa.com or fsca@oculus.de
- Quantity
- 19 systems with affected software
- Distribution
- U.S.: AL, AR, AZ, CAQ, CO, CT, FL, GA, HI, IA, ID, IL, KS, KY, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WY O.U.S.: N/A
- Lot, serial or model codes
- Software version: 1.26r26 and 1.26r27 Model Number: 70900 UDI-DI Code: 04049584000026 Serial Numbers: 70900 0574 8160 70900 0925 0240 70900 0995 2250 70900 1664 8130 70900 1995 2250 70900 3451 7060 70900 3795 2250 70900 3965 1270 70900 4041 7010 70900 4172 2110 70900 4795 2250 70900 6234 7150 70900 7552 1160 70900 7952 1180 70900 8804 6140 70900 8813 3140 70900 9142 0121 70900 9695 2250 70900 9795 2250
- 510(k) numbers
- ["K030719"]
- PMA numbers
- not on file
- Product code
- MXK
- Firm FEI
- 3002807782
- Firm city
- Wetzlar
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28