Siemens Healthcare Diagnostics, Inc.: IMMULITE 2000/IMMULITE 2000 XPi 3g Allergy specific IgE Universal Kit

Recall number
Z-1674-2018
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
IMMULITE 2000/IMMULITE 2000 XPi 3g Allergy specific IgE Universal Kit
Status
Terminated
Initiated
2018-01-02
Posted
2018-05-04
Terminated
2023-02-10
Reason
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Root cause
Error in labeling
Action
Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
Quantity
not on file
Distribution
Nationwide and Canada, Mexico, and Thailand
Lot, serial or model codes
3g Allergy specific IgE Universal kit, 3g Allergy specific IgE Universal kit (IMMULITE 2000/IMMULITE 2000 XPi L2KUN6 (D))  Lot codes: 3g Allergy Specific IgE Lot # 461 462 463 464 465 466 467 468 469 470 471 472 473 475 476 3g Allergy Specific IgE Lot # D463 D464 D465 D466 D467 D469 D471 D472 D476 D483
510(k) numbers
["K013134"]
PMA numbers
not on file
Product code
DHB
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register