Siemens Healthcare Diagnostics Inc: Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.

Recall number
Z-1673-2025
Recalling firm
Siemens Healthcare Diagnostics Inc
Product
Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.
Status
Open, Classified
Initiated
2025-03-31
Posted
2025-04-25
Terminated
not on file
Reason
Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.
Root cause
Under Investigation by firm
Action
An Urgent Medical Device Correction notice dated March 2025 was mailed to consignees. The notice instructed consignees to review the provided notification with their medical director to determine the appropriate course of action. Siemens will reimburse users for repeat tests and discarded units associated with this recall; consignees should contact the Customer Care Center for reimbursement. If product was further distributed, Siemens asks that the recall notification be forwarded. The recall notification is to be retained with laboratory records and forwarded to those who use the recalled devices. Consignees with any questions can contact Siemens Healthineers Technical Solutions Center at 977-229-3711.
Quantity
367 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, IL, KS, MI, MO, NJ, NY, OH, OK, PA, SC & WI. The countries of Austria, Canada, Chile, France, Germany, Iraq, Italy, Malaysia, North Macedonia, Saudi Arabia, Serbia, Slovenia, Spain, Sweden, Switzerland & Vietnam.
Lot, serial or model codes
Siemens Materials No. 10445071; UDI-DI: 0405686902174VK; Lot No. 234337002 and forward.
510(k) numbers
["K051650"]
PMA numbers
not on file
Product code
JHX
Firm FEI
3002637618
Firm city
Norwood
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register