FDA Medical Device Recalls (openFDA)
Ethicon Endo-Surgery Inc: HPBLUE: HARMONIC Blue Hand Piece; HP054: HARMONIC Hand Piece; Non-Sterile products are individually packed in a poly bag and then placed into an individual carton. Mfr. By: Ethicon Endo-Surgery, LLc, 475 Calle C, Guaynbo, PR 00969. Indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels.
- Recall number
- Z-1673-2012
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- HPBLUE: HARMONIC Blue Hand Piece; HP054: HARMONIC Hand Piece; Non-Sterile products are individually packed in a poly bag and then placed into an individual carton. Mfr. By: Ethicon Endo-Surgery, LLc, 475 Calle C, Guaynbo, PR 00969. Indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels.
- Status
- Terminated
- Initiated
- 2012-03-16
- Posted
- 2012-05-25
- Terminated
- 2013-03-05
- Reason
- Ethicon Endo-Surgery initiated a voluntary recall due to some lot numbers of the HARMONIC Hand Piece (HP054) and the HARMONIC Blue Hand Piece (HPBLUE), which impacts their compatibility with the HARMONIC Generator 300 (GEN04). The problem occurs after successful set up and initial use of HARMONIC Hand Pieces (HP054 and HPBLUE) with the Generator 300 (GEN04). The procedure is interrupted when
- Root cause
- Nonconforming Material/Component
- Action
- Ethicon Endo-Surgery Inc sent an URGENT: Product Correction Notification letter dated March 2012 to all affected customers. The letter identified the product, the problem, the action to be taken by the customer. Labels were enclosed that are used to identify the recalled product. Customers were instructed to fill out the Business Reply Form and fax it to Stericycle at 1-888-912-7091 within three business days for a credit. For questions regarding this recall call 513-337-3419.
- Quantity
- 1896 instruments
- Distribution
- Worldwide Distribution-USA including: AK,AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NCND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, and WY. and the countries of: Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Ecuador, Egypt, Greece, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Korea (South),Kuwait, Lebanon, Malaysia, Mexico, Netherlands, Antilles, Pakistan, Panama, Puerto Rico, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Taiwan, Thailand, Turkey, United Arab Emirates Uruguay, Venezuela & Yemen.
- Lot, serial or model codes
- HARMONIC¿ HP054 Hand Pieces Lot Numbers produced to-date: H92Y3M, J9039C, J90817, J90E7C, J9000C, J90432, J9084W, J90E7Z, J90029, J90470, J9088L, J90F3G,J9006C, J9054E, J9093L, J90F7Z, J9011D, J9056K, J9095C, J90G32, J90151, J9057V,J90974, J90G9Y,J9018H, J9058L, J90A45, J90H37,J9020F, J90624, J90A72, J90H4U, 9023D, J90681, J90C7X, J90H7D,J9027C, J9069F, J90C84, J90H9Y,J90314, J90726, J90D4N, J90J2N,J9034N, J90762, J90E0R, J90J8E,J9038W, J9076A, and J90E29. HARMONIC¿ HPBLUE Hand Pieces Lot Numbers produced to-date: J9001P, J9053N, J9091G, J90D84,J9005J, J9057E, J90A2C, J90E55,J9010E, J9065U, J90A64, J90F76,J9042G, J90696, J90C36, J90G4C,J9050E, J9086W, J90C7U, and J90H46.
- 510(k) numbers
- ["K063192"]
- PMA numbers
- not on file
- Product code
- LFL
- Firm FEI
- 1527736
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28