Skip to the content

FDA Medical Device Recalls (openFDA)

Oculus Optikgeraete GMBH: Pentacam AXL Wave REF 70020. Used to image the anterior segment of the eye.

Recall number
Z-1672-2022
Recalling firm
Oculus Optikgeraete GMBH
Product
Pentacam AXL Wave REF 70020. Used to image the anterior segment of the eye.
Status
Open, Classified
Initiated
2022-07-08
Posted
not on file
Terminated
not on file
Reason
Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs
Root cause
Software design
Action
On 07/08/2022, OCULUS GmbH issued an "Urgent safety information Corrective Action " Letter to its subsidiary OCULUS Inc via UPS express which identifies a software error in the software version 1.26r26 and 1.26r27 in the IOL calculator of the Pentacam software. Corrective measures to be taken: A software update to version 1.26r28 must be performed on all computers used and affected in connection with the Pentacam and additional workstations. Users should refrain from using the printout of the IOL calculator for affected installations until after the software update has been carried out. For questions or support - contact OCULUS Inc Service Team at 425-670-9977 or email service@oculususa.com or fsca@oculus.de
Quantity
21 systems with affected software
Distribution
U.S.: AL, AR, AZ, CAQ, CO, CT, FL, GA, HI, IA, ID, IL, KS, KY, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WY O.U.S.: N/A
Lot, serial or model codes
Software version: 1.26r26 and 1.26r27 Model Number: 70020 UDI-DI Code: 04049584025357 Serial Numbers: 70020 0111 0201 70020 0211 0201 70020 0311 0211 70020 0711 1230 70020 1211 0201 70020 1321 1211 70020 2411 0211 70020 2911 1250 70020 3511 0211 70020 3801 0230 70020 4701 0210 70020 4811 1250 70020 5411 0211 70020 5701 0210 70020 6611 1230 70020 6621 1211 70020 7011 0201 70020 7511 1230 70020 8021 1290 70020 8701 0210 70020 9011 0201
510(k) numbers
["K201724"]
PMA numbers
not on file
Product code
MXK
Firm FEI
3002807782
Firm city
Wetzlar
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register