Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6); Unicondylar knee prothesis

Recall number
Z-1671-2023
Recalling firm
Smith & Nephew, Inc.
Product
Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6); Unicondylar knee prothesis
Status
Open, Classified
Initiated
2023-05-02
Posted
2023-05-31
Terminated
not on file
Reason
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Root cause
Under Investigation by firm
Action
Smith & Nephew issued an Urgent Medical Device Recall Notice to its consignees on 05/02/2023 via email and letter. The notice explained the problem and the risk and requested the following: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customer s account number and name in Section A. 2. Please provide a copy of the attached Physician Communication to the surgical staff. The firm is seeking the return of the device.
Quantity
1596 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
a) REF 1-10011-100, UDI/DI 00885556873083; b) REF 1-10011-200, UDI/DI 00885556873090; c) REF 1-10011-300, UDI/DI 00885556873106; ALL BATCHES
510(k) numbers
["K190439"]
PMA numbers
not on file
Product code
HSX
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register