Ab Sciex: AB Sciex Triple Quad 4500MD LC/MS/MS System, Part No. 5031257; and QTRAP 4500MD LC/MS/MS System, Part No. 5031231. Mass spectrometers for in vitro diagnostic use only.

Recall number
Z-1671-2016
Recalling firm
Ab Sciex
Product
AB Sciex Triple Quad 4500MD LC/MS/MS System, Part No. 5031257; and QTRAP 4500MD LC/MS/MS System, Part No. 5031231. Mass spectrometers for in vitro diagnostic use only.
Status
Terminated
Initiated
2016-03-25
Posted
not on file
Terminated
2016-05-23
Reason
Under certain conditions, mass spectrometers may report incorrect quantitative results.
Root cause
Under Investigation by firm
Action
AB SCIEX issued Urgent: Medical Device Correction letter, dated 3/25/16, advising user of the problem and advising not to use the system until corrected. The letter also discussed the risk to health, as well as how to recognize if the issue has occurred. Sciex Field Service Employee (FSE) will be sent to customer site to perform the field corrective action on the affected system. Questions regarding the letter should be directed to SCIEX at 1-289-982-2531.
Quantity
2 units
Distribution
Distributed to the state of NC.
Lot, serial or model codes
Serial Numbers:  Triple Quad 4500MD LC/MS/MS System BX21081601 BX21101601 BX21091601 BX21111601 BX21121602 BX21131602 BX21141602 BX21151602  QTRAP 4500MD LC/MS/MS System BW20251601 BW20241601 BW20261602
510(k) numbers
not on file
PMA numbers
not on file
Product code
DOP
Firm FEI
3008935545
Firm city
Framingham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register