Cardio Medical Products: Cardio Medical Products, Inc. Defibrillation/Cardioversion/Monophasic or Biphasic Pacing/ECG Electrode Rx Only Single Patient Use Only Manufactured for Cardio Medical Products 385 Franklin Avenue, Suite L Rockaway, New Jersey 07866 www.cardiomedicalproducts.com (800) 227-3633 (973) 586-9624 FAX Defibrillator

Recall number
Z-1669-2013
Recalling firm
Cardio Medical Products
Product
Cardio Medical Products, Inc. Defibrillation/Cardioversion/Monophasic or Biphasic Pacing/ECG Electrode Rx Only Single Patient Use Only Manufactured for Cardio Medical Products 385 Franklin Avenue, Suite L Rockaway, New Jersey 07866 www.cardiomedicalproducts.com (800) 227-3633 (973) 586-9624 FAX Defibrillator
Status
Terminated
Initiated
2013-04-26
Posted
2013-07-08
Terminated
2016-08-15
Reason
Covidien received customer reports of Arcing/Sparking on the defibrillation electrode leadwire which may affect Cardio Medical Products Inc. defibrillator electrodes Model Number O101.
Root cause
Process control
Action
Cardio Medical Products, Inc. issued an Urgent Medical Device Recall letter dated April 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and segregate any product with the affected lot numbers and return affected product according to the instructions for product credit by contacting the Customer Service group at 1-800-962-9888, option 1, then option 2, and reference the recall letter to request a Return Goods Authorizing (RGA). Customers were instructed to complete the enclosed Verification Form and return it to Covidien by fax to 800-895-6140, even if they did not have any of the affected lots. Customers were also instructed to report any issues to the Quality Assurance Group to ensure proper device reporting procedures are followed. Customers with questions were instructed to call 1-800-962-9888, option 8, then ext 2500. For questions regarding this recall call 973-586-9500.
Quantity
657 units
Distribution
Nationwide Distribution
Lot, serial or model codes
510 K K103159  Part number O101  Lot numbers Y111412-2 Y111912-17 Y020613-03
510(k) numbers
["K103159"]
PMA numbers
not on file
Product code
LDD
Firm FEI
2246340
Firm city
Rockaway
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register