Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: SIEMENS Axiom Luminos dRF Max, a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Recall number
Z-1668-2015
Recalling firm
Siemens Medical Solutions USA, Inc
Product
SIEMENS Axiom Luminos dRF Max, a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
Status
Terminated
Initiated
2015-04-13
Posted
2015-05-23
Terminated
2017-03-20
Reason
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
Root cause
Software design
Action
A safety advisory notice, dated April 13, 2015, was sent to end users with affected devices to alert them about the issue and associated risks,and to provide instructions for avoiding potential risk. A new software version will be released to permanently resolve the issue.
Quantity
133 total
Distribution
Nationwide Distribution.
Lot, serial or model codes
Model Number of device  10762471with serial numbers: 5043 5085 5069 5086 5037 5072 5049 5039 5120
510(k) numbers
["K062623"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register