Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: InSpan Inserter, Right Assembly; InSpan Inserter, Left Assembly Product Usage: Inserters are stainless steel instruments that have an inner shaft. The left and right inserters are used in combination with each other to place the implants into the body. Inserters are used to install the InSpan Spinous Plate System for the purpose of achieving supplemental fixation.

Recall number
Z-1666-2013
Recalling firm
SpineFrontier, Inc.
Product
InSpan Inserter, Right Assembly; InSpan Inserter, Left Assembly Product Usage: Inserters are stainless steel instruments that have an inner shaft. The left and right inserters are used in combination with each other to place the implants into the body. Inserters are used to install the InSpan Spinous Plate System for the purpose of achieving supplemental fixation.
Status
Terminated
Initiated
2013-05-07
Posted
2013-07-05
Terminated
2016-01-11
Reason
Potential for set screw to be come deformed.
Root cause
Device Design
Action
Spine Frontier sent an Urgent Advisory Notice letter dated May 30, 2013 to all affected consignees. The letter identified the affected product. problem and actions to be taken. The letter informed consignees that a correction was implemented for InSpan Inserters. Consignees were instructed to remove parts from the field immediately for modification and that replacement Inserters will be shipped upon receipt of completed acknowledgement form and RMA request for part return. For questions call 978-232-3990.
Quantity
61 individual inserter assemblies
Distribution
Worldwide Distribution - USA Nationwide: in the states of: FL, CO, VA, KS, TX, OH, TN, OK, CA, IL, MO, MN, Wash DC, including PR and the countries of Dominican Republic, Medico, Panama
Lot, serial or model codes
Right Assembly: 43097, 44614, 43097R; Left Assembly: 43098, 44615, 43098R
510(k) numbers
not on file
PMA numbers
not on file
Product code
HXX
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register