Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Molecular, Inc.: The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).

Recall number
Z-1665-2025
Recalling firm
Abbott Molecular, Inc.
Product
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
Status
Open, Classified
Initiated
2025-03-20
Posted
2025-04-24
Terminated
not on file
Reason
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Root cause
Device Design
Action
An Urgent Field Safety Notice/ Field Correction Recall notification letter was sent to customers. Necessary Actions In accordance with the Alinity m Operations Manual, users should always utilize Personal Protective Equipment when engaging with the instrument. In the event you observe a leak, please take appropriate precautions to prevent exposure and immediately contact your Abbott Molecular Representative for additional troubleshooting information. Please complete and return the associated Customer Reply Form. If you have forwarded this product to any other laboratories, please also forward this letter and customer reply form to that laboratory. Review this information with laboratory personnel and retain this communication for future reference. If you have any questions regarding this communication, please contact your local Abbott representative. We apologize for any inconvenience this may have created for your laboratory.
Quantity
not on file
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
All Serial Numbers. 08N53-001, UDI-DI: 00884999047389; 08N53-002, UDI-DI: 00884999048034; 08N53-032, UDI-DI: 00884999047587
510(k) numbers
not on file
PMA numbers
not on file
Product code
OOI
Firm FEI
3005248192
Firm city
Des Plaines
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register