FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy
- Recall number
- Z-1665-2023
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy
- Status
- Open, Classified
- Initiated
- 2023-05-05
- Posted
- 2023-05-30
- Terminated
- not on file
- Reason
- A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601
- Root cause
- Under Investigation by firm
- Action
- Elekta notified customers, via email, on 05/05/2023. Customers were instructed that Elekta will contact each impacted customer to perform an onsite inspection of the power supply unit and replace as needed. Customers were asked to complete and return the Acknowledgement Form.
- Quantity
- 46 units
- Distribution
- Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA. International distribution to Belgium, Canada, Chile, China, France, Germany, India, Iraq, Italy, Jordan, Libyan Arab Jamahiriya, Mauritius, Mexico, Morocco, Poland, Russian Federation, Spain.
- Lot, serial or model codes
- Varex Part Number 95510354H-01 (Varex serial numbers 354-01H_A0010004PC to 354-01H_A0010269PC). a. Elekta Infinity: UDI-DI Number: 05060191071543, 05060191071512 (Serial Numbers: 156073, 152223, 153681); b. Versa HD: UDI-DI Number: 05060191071574, 05060191071529 (Serial Numbers: 156941, 156989, 156724, 156998); c. Elekta Synergy: UDI-DI Number: 05060191071550, 05060191071505 (Serial Number: 152483) Additional serial numbers distributed OUS: 109331, 109333, 109335, 109336, 109339, 109341, 109342, 151879, 152194, 153038, 153299, 153410, 153970, 156841, 156849, 156966, 156968, 156969, 156970, 156971, 156973, 156977, 156978, 156979, 156980, 156981, 156982, 156983, 156984, 156985, 156986, 156987, 156988, 156990, 156992, 156995, 156996, 156997.
- 510(k) numbers
- ["K210500"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28