FDA Medical Device Recalls (openFDA)
SpineFrontier, Inc.: S-LIFT Inserter Outer Sleeve Product Usage: The Inserter Sleeve is an optional instrument to add further compression of the Inserter into the Implant during implantation of S-LIF Intervertebral Body Fusion Devices using bone graft to facilitate fusion of the spine of skeletally mature patients. Model: SI50003 RevA
- Recall number
- Z-1665-2013
- Recalling firm
- SpineFrontier, Inc.
- Product
- S-LIFT Inserter Outer Sleeve Product Usage: The Inserter Sleeve is an optional instrument to add further compression of the Inserter into the Implant during implantation of S-LIF Intervertebral Body Fusion Devices using bone graft to facilitate fusion of the spine of skeletally mature patients. Model: SI50003 RevA
- Status
- Terminated
- Initiated
- 2013-05-24
- Posted
- 2013-07-05
- Terminated
- 2014-10-08
- Reason
- The outer sleeve would not fit over the S-LIFT Inserter assembly because the hole at the top of the sleeve handle was too small.
- Root cause
- Device Design
- Action
- Spine Frontier sent an Urgent Advisory Notice letter dated On June 7, 2013 to all consignees via email. The letter identified the affected product, problem and actions to be taken. The letter informed consignees that a modification has been implemented to the Outer Sleeve component of the S-LIFT Inserter and that SpineFrontier has initiated removal of the affected product from the field.. An acknowledgement form is to be signed and returned from each recipient of the Advisory Notice. Upon receipt of the acknowledgement form and RMA request for part return, replacement parts will be shipped. For questions call 978-232-3990.
- Quantity
- 12 devices
- Distribution
- USA Nationwide Distribution including the states of: FL, TX, OK, MO, NJ, MA
- Lot, serial or model codes
- Lot/Serial Numbers: 47175
- 510(k) numbers
- ["K092815"]
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3005977257
- Firm city
- Beverly
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28