Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: TSX-301A: AQUILION ONE System AQUILION ONE is a multislice CT system that supports whole body scanning. The system supports simultaneous data acquisitions of 320 slices using the Selectable Slice-thickness Multi-row Detector.

Recall number
Z-1665-2011
Recalling firm
Toshiba American Medical Systems Inc
Product
TSX-301A: AQUILION ONE System AQUILION ONE is a multislice CT system that supports whole body scanning. The system supports simultaneous data acquisitions of 320 slices using the Selectable Slice-thickness Multi-row Detector.
Status
Terminated
Initiated
2010-07-22
Posted
2011-03-16
Terminated
2011-04-05
Reason
Toshiba America Medical Systems (TAMS) has found that communication may be cut off in the Aquilion ONE/ Premium Systems, owing to the failure of a part used in the console.
Root cause
Software design
Action
The firm, Toshiba America Medical Systems (TAMS), sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated July 22, 2010, to its customers. The letter described the product, problem and actions to be taken. The letter informed the customers that a new relay will be installed to prevent the occurrence of the problem within the next 90 days and that their Toshiba Service Representative will contact them to install the part to their Aquilion ONE/Premium System. The customers were instructed to contact their Toshiba Service Representative, if the problem occurs before the corrective action has been taken; to share the information with all users and review radiologists as well as clinical engineering or Biomedical group at their facility, and to complete and return the attached CUSTOMER REPLY FORM via fax to 877-349-3054 or email to raffairs@tams.com. If you have any questions regarding this letter, please feel free to contact the Director, Regulator Affairs at (800) 421-1968 or your local Toshiba Representative at (800) 521-1968.
Quantity
43 units
Distribution
Nationwide distribution: USA including states of: AL, AR, AZ, CA, CO, FL, GA, IA, KY, LA, MA, MD, MI, MN, MT, ND, NJ, NV, NY, OH, PA, PR, TX, VA, WI, and WY.
Lot, serial or model codes
Last 4 of serial number: 2005, 2008, 2010, 2023, 2026, 2034, 2036, 2038, 2048, 2049, 2051, 2052, 2053, 2061, 2063, 2067, 2070, 2071, 2072, 2077, 2080, 2092, 2115, 2116, 2006, 2007, 2008, 2123, 2131, 2133, 2011, 2139, 2143, 2148, 2151, 2014, 2161, 2162, 2165, 2167, 2168, 2169, 2170, 2002, 2004, 2158, 2166, 2174
510(k) numbers
["K083282"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register