FDA Medical Device Recalls (openFDA)
Beckman Coulter, Inc.: Beckman Coulter DxI 9000 Access Immunoassay Analyzer
- Recall number
- Z-1664-2025
- Recalling firm
- Beckman Coulter, Inc.
- Product
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer
- Status
- Open, Classified
- Initiated
- 2025-03-27
- Posted
- 2025-04-24
- Terminated
- not on file
- Reason
- When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
- Root cause
- Software design
- Action
- Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/27/2025 via USPS first class mail. The notice explained the issue with the device, impact to the patient, and provided a workaround until a software upgrade is available. If the consignee further distributed the affected device, they were directed to provide a copy of the notice to the recipient. For questions, contact Customer Support Center, from the website: http://www.beckmancoulter.com
- Quantity
- 370 units
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- UDI/DI 15099590732103, all units run with system software versions SW 1.20.0 and below
- 510(k) numbers
- ["K221225"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2122870
- Firm city
- Chaska
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28