Skip to the content

FDA Medical Device Recalls (openFDA)

BD KIESTRA LAB AUTOMATION: BD Kiestra" ReadA; Catalog No.: 446948.

Recall number
Z-1658-2026
Recalling firm
BD KIESTRA LAB AUTOMATION
Product
BD Kiestra" ReadA; Catalog No.: 446948.
Status
Open, Classified
Initiated
2026-02-06
Posted
2026-03-27
Terminated
not on file
Reason
In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.
Root cause
Software Manufacturing/Software Deployment
Action
On February 6, 2026 customers were contacted via phone call with a follow-up URGENT: Medical Device Correction letter were sent to customers on or about March 10, 2026. Actions to be taken: 1. Ensure the contents of this notification are read and understood. 2. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088)
Quantity
82 units (7 US, 65 OUS)
Distribution
Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;
Lot, serial or model codes
Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All;
510(k) numbers
not on file
PMA numbers
not on file
Product code
JTQ
Firm FEI
3010141591
Firm city
Drachten
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register