FDA Medical Device Recalls (openFDA)
Biomet, Inc.: Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush onto the natural glenoid surface.
- Recall number
- Z-1658-2008
- Recalling firm
- Biomet, Inc.
- Product
- Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush onto the natural glenoid surface.
- Status
- Terminated
- Initiated
- 2008-03-13
- Posted
- 2008-09-16
- Terminated
- 2008-10-15
- Reason
- The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien
- Root cause
- Device Design
- Action
- Consignees were notified by Urgent Medical Device Removal Notice dated 3/13/08 to discontinue use of the instruments and return to the firm. Operating room managers were sent a separate letter dated 3/13/08 and informed of the possible effects to patients from previous use of this instrument and it was suggested they monitor patients for these effects. If you have any questions contact Mary Hardesty at 1-800-348-9500 or (574) 372-3983.
- Quantity
- 189
- Distribution
- Class II Recall - Nationwide Distribution.
- Lot, serial or model codes
- Lots 44550, 110100, 171760, 211530, 211550, 767940, 820310, 820330, 827250, 853810, 866590, 919780, 919790, 919800, 968270, 980230 and 980250.
- 510(k) numbers
- ["K060694"]
- PMA numbers
- not on file
- Product code
- KWT
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28