Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush onto the natural glenoid surface.

Recall number
Z-1658-2008
Recalling firm
Biomet, Inc.
Product
Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush onto the natural glenoid surface.
Status
Terminated
Initiated
2008-03-13
Posted
2008-09-16
Terminated
2008-10-15
Reason
The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien
Root cause
Device Design
Action
Consignees were notified by Urgent Medical Device Removal Notice dated 3/13/08 to discontinue use of the instruments and return to the firm. Operating room managers were sent a separate letter dated 3/13/08 and informed of the possible effects to patients from previous use of this instrument and it was suggested they monitor patients for these effects. If you have any questions contact Mary Hardesty at 1-800-348-9500 or (574) 372-3983.
Quantity
189
Distribution
Class II Recall - Nationwide Distribution.
Lot, serial or model codes
Lots 44550, 110100, 171760, 211530, 211550, 767940, 820310, 820330, 827250, 853810, 866590, 919780, 919790, 919800, 968270, 980230 and 980250.
510(k) numbers
["K060694"]
PMA numbers
not on file
Product code
KWT
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register